Error messages – KERN & SOHN HFB 600K200 User Manual
Page 26
HFB-BA-e-1212
26
8. Error messages
Error
message
Description
Possible causes
--ol-
Maximum load exceeded
Reduce load
Check whether the balance has
been damaged
Err 4
Zeroing range exceeded
(normally 4% max)
Overload when zeroing
Err 5
Keyboard error
Improper operation of the balance
Err 6
Value outside the A/D changer
range
Weighing plate not installed
Damaged weighing cell
Damaged electronics
Err 9
At an instable weighing value
ZERO or TARE button
actuated
Wait for a stable weighing value
Err 17
Tare value out of range
Reduce tare load
Ba lo
Capacity of rechargeable
battery exhausted
Recharge battery
Should other error messages occur, switch balance off and then on again. If the error