5.7.1 Data-Hold function
Switch on the balance, keep the
HOLD-key pressed until the current setting „Hx“
(H1 – H6) appears.
Press
ON/OFF button repeatedly until the desired setting „H1-H4“ is displayed.
Confirm setting with the
HOLD button.
Suspend the material to be weighed
The weight value is fixed and displayed depending on the setting (H1 – H4) (see
chap. 5.7), represented by the [
] in the upper left.
5.7.2 Animal weighing function
This function is suitable for busy weighing procedures. The result is a mean value
formed by 16 weighing values which is found out within 3 seconds.
Switch on the balance, keep the
HOLD-key pressed until the current setting „Hx“
(H1 – H6) appears.
Press
ON/OFF button repeatedly until the setting „H5“ is displayed.
Confirm setting with the
HOLD button.
Suspend the material to be weighed
Press the
HOLD-key, the display counts reverse from 3 -1.
The calculated mean value is indicated, symbolised by the [
] top left.
Before further measurements press first the
ON/OFF/TARE-key.