Briefly press the, Key; the right digit starts flashing. press the – KERN & SOHN VHS 2T1 Addendum User Manual
Page 4
![background image](/manuals/697337/4/background.png)
Briefly press the
TARE
key; the right digit starts flashing.
Press
the
PRE-
TARE
,
TOTAL
and
TARE
function keys to enter the medium adjustment
value (e.g. 1250 kg) as described above, press the
TARE
key to confirm all
segments until only the NET display arrow flashes.
Import the adjustment weight (in this case the 1250 kg above).
To confirm press the
TARE
key for c. 4 sec. The display reverts to “AF00“. The
display shows again the medium adjustment value; the medium adjustment
load (1250 kg) can now be taken off.
Press
the
TOTAL
key, the NET + ~ display arrows are flashing, the last
adjustment value can be entered.
Briefly press the
TARE
key; the right digit starts flashing.
Press the function keys
PRE-
TARE
,
TOTAL
and
TARE
to enter the last adjustment value
(e.g. 1875 kg) as described above, press the
TARE
key to confirm all segments
until only the NET +
~ display arrows flash.
Import the adjustment weight (in this case the 1875 kg above).
To confirm press the
TARE
key for c. 4 sec. The display reverts to “AF00“. Again
the display shows the last adjustment value.
To exit adjustment mode, press the
TOTAL
key; the scales show e. g. AP 11
(internal value)
Press
the
TARE
key for c. 4 sec; the system reverts to weighing mode, the
scales show the current weight, the adjustment is complete, the adjustment
load (1875 kg) may now be taken off.
Please note: If you readjust adjusted scales with the help of multi-point adjustment,
you should use the smallest weight first, followed by the medium weight and finally
the greatest weight. Use only 1 adjustment value, then set value 2 and 3 to "0000.0“.
KERN VHB_VHS