8.5 Instant help
In case of an error in the program process, briefly turn off the weighing system and
disconnect from power supply. The weighing process must then be restarted from
the beginning.
Help:
Fault
Possible cause
The displayed weight is permanently
changing
Draught/air movement
Floor vibrations
Weighing beams come into contact with
foreign substances.
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing result is obviously
incorrect
No zero display for unloaded weighing
beams.
Adjustment is no longer correct.
Great fluctuations in temperature.
The weighing beams are not level.
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.