10.1 Overview, setting regulation, tolerances
Testing and setting regulations:
Capacity
600 kg
1500 kg
Readability
200 g
500 g
Min
4 kg
10 kg
Max
600 kg
1500 kg
1/3 corner load
200 kg
500 kg
Tolerance
200 g
500 g