5 instant help, Instant help – KERN & SOHN BAN 3T-3M User Manual
Page 12

8.5 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible
cause
• Draught/air movement
• Floor vibrations
The displayed weight is permanently
changing
• Weighing plate has contact with other
objects.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
• No zero display with unloaded balance
• Adjustment is no longer correct.
The weighing result is obviously
incorrect
• Great fluctuations in temperature.
• Weighing bridge on an uneven surface.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
KFP V40-IA-e-1110
12