5.7.5 Position adjust switch
FOB-BA-e-1118
15
Adjust switch at the lower surface
(depending on the model, on the right
or
the
left
side)
5.8
Verification
General:
According to the EU guideline 90/384/EEC balances must be verified if they are to be
used as follows (legally regulated area):
a) For commercial transactions if the price of goods is determined by weighing
b) For the production of medines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory
c) For official purposes
d) For the production of finished packages
In case of doubt, please contact your local office of weights and measures.
Verification information
An EU qualification approval is available for those balances marked as appropriate
for verification in the technical data. In the event that the balance is applied in an
area subject to verification as described above, it must be verified and re-verified at
regular intervals.
Re-verification of a balance is carried out in compliance with the respective legal pro-
visions of the states. The term of verification validity for balances in Germany, for
example, is normally 2 years.
The legal provisions of the country of use are to be observed.