5.4 Batterie operation

5.5 Connection of peripheral devices

Before connecting or disconnecting of additional devices (printer, PC) to the data interface,
always disconnect the balance from the power supply. With your balance, only use
accessories and peripheral devices by KERN, as they are ideally tuned to your balance.

5.6 Initial Commissioning

A warming up time of 2 hours after switching on stabilizes the measuring values.
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter “Adjustment”.

5.7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each
balance must be coordinated - in compliance with the underlying physical weighing principle
- to the existing acceleration due to gravity at its place of location (only if the balance has
not already been adjusted to the location in the factory). This adjustment process must be
carried out during the initial start-up, after change in location and variation of surrounding
temperature. To receive accurate measuring values it is also recommended to adjust the
balance periodically in weighing operation.