beautypg.com

KERN & SOHN IKT User Manual

Page 15

background image

KERN

IKT 6K1M IKT 12K2M IKT 30K5M

IKT

60K10LM

IKT

120K20LM

Accuracy class

III

III

III

III

III

Readability (d)

1 g

2 g

5 g

10 g

20 g

Verification value (e)

1 g

2 g

5 g

10 g

20 g

Weighing range (max)

6 000 g

12 000 g

30 000 g

60 000 g

120 000 g

Minimum load (Min)

20 g

40 g

100 g

200 g

400 g

Taring range
(subtractive)

6 000 g

12 000 g

30 000 g

60 000 g

120 000 g

Reproducibility

0.5 g

1 g

2.5 g

5 g

10 g

Linearity

±0.5 g

±1 g

± 2.5 g

± 5 g

± 10 g

Smallest piece weight

1 g

2 g

5 g

10 g

20 g

Adjustment points

2/5/6 kg

2/5/10/12 kg 10/20/30 kg 20/50/60 kg

20/50/100/

120 kg

Recommended
adjusting weight M1
(not supplied)

10 kg

10 kg

20 + 10 kg

50 kg

50kg+50kg

Humidity of air

max. 80% rel. (non-condensing)

Stabilization time
(typical)

3 sec.

Allowable ambient
temperature

+10 °C … + 40 °C

Warm-up time

2 h

2 h

2 h

2 h

2 h

Housing (B x D x H)
mm

315 x 305 x 70

450 x 350 x 115

Vibration filter

yes

Weighing plate
stainless steel mm

315 x 305

450 x 350

Units

see menu

Weight kg (net)

7.5

9.5

Data interface

yes (RS232)

Master data storage for
all operating modes

40

Print forms with up to
20 informations

1

Recipes with up to 7
components

5

15 FKT_IKT_PKT-BA-e-1435