9.4 Interface parameters
Data output is carried out via interface RS 232 C.
General
The previous condition for the data transfer between balance and a peripherical
device (e.g. printer, PC ...) is that the appliances are set to the same interface
parameters (e.g. baud rate, transfer mode ...).
Data transfer
mode
0.0
g
Vn�
T
PR�
In weighing mode keep the PRINT key pressed until [Unit]
appears.
Press the MODE button several times until „Pr“ is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired settings by pressing the MODE key
rE CR
Data output via remote control commands
Pr PC
Data output using the PRINT key
AU PC Continuous data output
bA Pr
Output on bar code printer
AU Pr
Autom. data output of stable weighing values
Use the SET key to confirm selection. The balance returns to
weighing mode.
FKB_A-BA-e-1513