Flintec DAD 141.1 Modbus Communication User Manual
Page 8
![background image](/manuals/676052/8/background.png)
DAD 141.1 Technical Manual Modbus Communication, Rev. 2.3 - Juni 2014
Page
8
The values of the update rate are 0 to 7 (2 exp 0 = 1, 2 exp 7 = 128).
see also command description:
UR Update Rate
2122
Int32
2
RW
Zero Tracking
This Index enables or disables the zero tracking (TAC protected).
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 2122.
The values of Zero Tracking function are 0 to 255.
see also command description:
ZT Zero Tracking
2128
Int32
2
RW
Analog Output
Mode
This Index is used for set the analog output mode.
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 2128.
The values of the output mode are 0 to 5
see also command description:
AM Analog Output
Mode
2200
Int32
2
RW
Absolute Gain
Calibration
This Index Reads/Modify the absolute gain point (TAC protected).
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 2200.
see also command description:
AG Absolute Gain
2202
Int32
2
RW
Absolute Zero
Calibration
This Index Reads/Modify the absolute zero point (TAC protected).
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 2202.
see also command description:
AZ Absolute Zero
2204
Int32
2
RW
Calibrate Enable
This Index sets the calibration functions to the enabled state.
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 2204.
see also command description:
CE Calibrate Enable
2206
Int32
2
RW
Calibrate Gain
This Index sets the calibration gain (span) value (TAC protected).
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 2206.
see also command description:
CG Calibrate Gain
220C
Int32
2
RW
Calibrate Max
This Index Reads/Modify the maximum allowable output value (TAC protected).
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 220C.
see also command description:
CM Maximum Output
220E
Int32
2
RW
Calibrate Min
This Index Reads/Modify the minimum allowable output value (TAC protected).
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 220E.
see also command description:
CI Minimum Output
2212
Int32
2
W
Calibrate Zero
This Index sets the calibration zero point (TAC protected).
The 32 bit integer data accessed by writing 2 16-bit registers from index 2212.
see also command description:
CZ Calibrate Zero
2214
Int32
2
RW
Decimal Point
This Index Reads/Modify the decimal point position (TAC protected).
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 2214.
see also command description:
DP Decimal Point
Index (hex)
Type
Size Access Function
Description
2216
Int32
2
RW
Display Step
Size
This Index Reads/Modify the the display step size (TAC protected).
The 32 bit integer data accessed by reading or writing 2 16-bit registers from index 2216.
see also command description:
DS Display Step Size