Up dwn set esc, 4ma 20ma exit, Up dwn set – Eagle Microsystems RA1000 User Manual
Page 17
17
0.00ppm
UP DWN SET
ESC
1
UP
Increase CL
2
value
2
DWN
Decrease CL
2
value
3
SET
The value entered is the CL
2
residual at the time the STR pushbutton was pressed.
4
ESC
Go back to the get sample screen
CALIBRATING THE ANALOG OUTPUT
For one to calibrate the analog output, an accurate DMM or equivalent current meter (20mA
range) must be connected alone or in series with the 4 to 20mA (PO1) load.
SET ANALOG OUT
4mA 20mA
EXIT
1
4mA
Set 4mA point
2
20mA
Set 20mA point
3
No function
4
EXIT
Go back to the calibrate menu.
SET XX.00mA OUT
UP DWN
SET
The XX.00 is either 4.00 or 20.00 and does not change when setting the selected point.
Observe the current reading from the DMM and use the up and down pushbuttons to set the
selected point to 4/20mA.
1
UP
Increase current
2
DWN
Decrease current
3
No function
4
SET
Store new 4mA or 20mA point