CALIBRATING THE ANALOG OUTPUT

For one to calibrate the analog output, an accurate DMM or equivalent current meter (20mA

range) must be connected alone or in series with the 4 to 20mA (PO1) load.

SET ANALOG OUT

4mA 20mA

EXIT

1

4mA

Set 4mA point

2

20mA

Set 20mA point

3

No function

4

EXIT

Go back to the calibrate menu.

SET XX.00mA OUT

UP DWN

SET

The XX.00 is either 4.00 or 20.00 and does not change when setting the selected point.

Observe the current reading from the DMM and use the up and down pushbuttons to set the
selected point to 4/20mA.

1

UP

Increase current

2

DWN

Decrease current

3

No function

4

SET

Store new 4mA or 20mA point