Defibtech DDU-2450 Series Operating Guide User Manual
Page 9

DAC-E2530EN-BC
DAC-E2530EN-BC
16
17
For more detailed information, refer to the User Manual (on Defibtech User CD).
MAINTENANCE MODE
Maintenance Mode permits the user to perform
maintenance-related actions such as viewing unit
information, initiating unit self-tests, changing
unit parameters, downloading rescue data, and
upgrading software.
To enter Maintenance Mode press
the bottom softkey button to the right
of the Tool Icon (shown at left) on
the AED Status screen (for instructions on how to
access the AED Status screen, see previous page).
To exit Maintenance Mode, simply turn the unit off
by pressing the ON/OFF button. For comprehensive
information about Maintenance Mode, refer to
Chapter 8 of the User Manual (on Defibtech User CD).
CHECKING AED STATUS
ROUTINE MAINTENANCE
•
Flashing Green: The DDU-2000 Series AED is OFF and
ready for use.
•
Solid Green: The DDU-2000 Series AED is ON and ready
for use.
•
Flashing or Solid Red: The DDU-2000 Series AED needs
immediate service. Refer to the“Troubleshooting” section
of the User Manual (on Defibtech User CD) or call Defibtech
for service.
•
No Flashing Light: The DDU-2000 Series AED needs
immediate service. Refer to the “Troubleshooting” section
of the User Manual (on Defibtech User CD) or call Defibtech
for service.
Active
Status
Indicator
Although the DDU-2000 Series AED is designed to be very low maintenance, simple mainte-
nance tasks must be performed by the owner/operator on a regular basis to ensure the unit’s
dependability.
Daily Monthly After Each Use Action
•
•
•
Check that the Active Status Indicator is flashing green
•
•
Check the condition of the unit and accessories
•
Run manually initiated self-test
•
Replace pads
•
Check pad and battery pack expiration dates
•
Check the DDC card, if one was installed
If the unit needs attention, refer to the “Troubleshooting” section of the User Manual
(on Defibtech User CD) or call Defibtech for service. For contact information, refer to the
“Contacts” section of this guide.
ACTIVE STATUS INDICATOR (ASI)
Visually check the Active Status Indicator (ASI)
on a daily basis. The ASI should flash green.
If the ASI flashes red, is solid red, or if there is no flashing light, the unit requires service.
Anytime the ASI flashes red, the unit will also “beep” periodically to call attention to itself.
The display screen during
Maintenance Mode
AED STATUS SCREEN
The AED Status screen is used to provide
a quick overview of the DDU-2000 Series
AED’s status and to display select informa-
tion without turning the unit on for a rescue.
With the AED off, press and release the
CENTER softkey button to display the AED
Status screen. The AED Status screen will
be displayed for a short period of time.
If the unit does not turn on at all, check to
make sure a good battery pack is installed.
(Refer to the “Troubleshooting” section of
the User Manual (on Defibtech User CD).
AED Status screen
AED status
OK
Battery status
Expires 01/2018
Pads status
Adult
Expires 06/2018
AED S/N
099999999
Battery S/N
099999999
Software version
V2.2
Rescue now
AED status
AED maintenance
AED options
Rescue options
Help topics
Turn AED off
AED Main Menu