12. ADJUSTING HEIGHT OF THE TIP EJECTOR
A height of the tip ejector can be 
adjusted so that it can match the 
conditions of how the tip used was 
connected.
Use a small minus screwdriver. 
By turning the adjustment screw in 
a counter-clockwise direction, the 
tip ejector can be lowered toward 
the tip side. 
13. STORAGE AND MAINTENANCE
When cleaning the outside of the pipette or the tip holder, remove dirt by using a cloth 
dampened with 60% isopropyl, 70% ethanol or ph-neutral detergent, and wipe it using 
a lint-free and dry soft cloth. Also, because the ejector can easily be removed, the tip 
holder can easily be cleaned. 
 
 
 
 
 
 
 
 
 
 
 
Regularly clean the tip holder about once every half a year to a year, depending on 
the frequency used. Regularly confirm the pipette accuracy. 
Maintenance
When inspecting the pipette or managing pipette precision, the following device can 
be used. (Refer to “17. LIST OF ITEMS SOLD SEPARATELY ( DISPOSABLE 
ITEMS)”) 
- Leak tester (Sold separately: AD-1690):
Leakage within the pipette can be easily checked.
- Pipette accuracy tester (Sold separately: BM series, AD-4212B-PT, FX-300i-PT):
By using an electronic balance and specified software “WinCT-pipette”, the actual 
dispensed amount of the pipette can be confirmed.
Expanded
Ejector
Tip holder
Adjustment screw