7 blowout function – A&D Weighing MPA-1200 User Manual
Page 29

27
2) Operating the reverse mode
 
 
[ 1 ] Press the Operation key on the pipette to 
aspirate the liquid.
 
[ 2 ] Press the Operation key on the pipette to 
dispense the liquid.
 
 
 
 
 
 
[ 3 ]
Press the Operation key on the pipette to
discharge the remaining liquid.
 
 
 
 
 
 
9-7 Blowout function
This is the function to forcibly dispensing the liquid remaining in the end of the tip by 
temporarily lowering the piston built in the pipette below the start position for aspiration 
after dispensing the liquid remaining in the tip. 
By pressing the Operation key when “bL” is shown on the display, carry out 
blowout. 
* After carrying out blowout, the built-in piston remains in the blowout position while the 
 Operation key is being held down, and it returns to initial position when the 
finger was released from the Operation key. By releasing the Operation key after 
removing the tip end from the vessel, aspiration of the dispensed liquid in 
the tip again can be prevented. 
Aspirate
Dispense
Discharge the
remaining
liquid