A&D Weighing MPA-1200 User Manual
Page 34

32
With theμL unit selected, the Notice
mark is turned off while the μL
unit mark lights up.
With the mg unit selected, the Notice mark
blinks while the μ L unit mark is
turned off
Method for selecting the mg unit
[ 1 ] Press the Back key to enter the
mode to select an operation mode.
[ 2 ] Press the Up or Down key to
select the system setting mode
“SYS”.
[ 3 ] Press and hold the Enter key for
approx. three seconds.
[ 4 ] The unit currently selected is displayed.
* The calibration data reverts back to the factory default once the weight unit is
switched.
11-4 Weight calibration function (mg calibration function)
The density of a liquid varies depending on the type and concentration of the material.
Make sure to perform mg calibration when you dispense a different sample or use the
mg unit for the first time. Further, when you need to always control dispensed amounts
in a precise manner, perform mg calibration when you change the dispensing amount
setting as well.
[ 5 ] Press the Up or Down
key to select the unit.
[ 6 ] Press Enter key to
confirm.
(Unit in μL)
mg
(Unit in mg)