Title block and end block, Operation – A&D Weighing FX-3000iWP User Manual
Page 54
52
Title block and end block
When a weight value is recorded as the GLP data, “Title block” and “End block” are inserted at the
beginning and at the end of a group of weight values, in the GLP report.
Note
To output the report to an AD-8121B , use MODE 3 of the AD-8121B.
Operation
1 With the weight data displayed, press and hold the
key until S5tart is displayed.
The “Title block” is output.
2 The weight data is output according to the parameter setting of the data output mode.
3 Press and hold the
key until recend is displayed. The “End block” is output.
When the setting is “info 1”:
When the setting is “info 2”:
Weight data
Time
Signature
End block
AD-8121 format
Manufacturer
Model
Serial number
ID number
Date
Time
Title block
General data format
A & D
MODEL FX-3000i
S/N 01234567
ID ABCDEFG
DATE
START
TIME
WT +1234.56 g
WT +1234.61 g
WT +1234.62 g
WT +1234.63 g
WT +1234.51 g
WT +1234.53 g
WT +1234.71 g
WT +1234.64 g
END
TIME
SIGNATURE
----------------
A & D
MODEL FX-3000i
S/N 01234567
ID ABCDEFG
DATE 2006/02/06
START
TIME 12:23:34
WT +1234.56 g
WT +1234.61 g
WT +1234.62 g
WT +1234.61 g
WT +1234.61 g
WT +1234.53 g
WT +1234.71 g
WT +1234.64 g
END
TIME 12:34:56
SIGNATURE
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
CR
CR
LF
CR LF