A&D Weighing FX-3000i User Manual
Page 52
50
Title block and end block
When a weight value is recorded as the GLP data, “Title block” and “End block” are inserted at the
beginning and at the end of a group of weight values, in the GLP report.
Note
To output the report to an AD-8121B , use MODE 3 of the AD-8121B.
Operation
1 With the weight data displayed, press and hold the PRINT key until S5tart is displayed.
The “Title block” is output.
2 The weight data is output according to the parameter setting of the data output mode.
3 Press and hold the PRINT key until recend is displayed. The “End block” is output.
When the setting is “info 1”:
When the setting is “info 2”:
Weight data
Time
Signature
End block
AD-8121 format
Manufacturer
Model
Serial number
ID number
Date
Time
Title block
General data format
A & D
MO DE L FX -3 00 0i
S/ N 01 23 45 67
ID A BC DE FG
DA TE
ST AR T
TI ME
WT + 1234 .5 6 g
WT + 1234 .6 1 g
WT + 1234 .6 2 g
WT + 1234 .6 3 g
WT + 1234 .5 1 g
WT + 1234 .5 3 g
WT + 1234 .7 1 g
WT + 1234 .6 4 g
EN D
TI ME
SI GN AT URE
- - -- - - - - - - - - - - - -
A & D
MODEL FX-3000i
S/N 01234567
ID ABCDEFG
DATE 2008/12/31
START
TIME 12:23:34
WT +1234.56 g
WT +1234.61 g
WT +1234.62 g
WT +1234.61 g
WT +1234.61 g
WT +1234.53 g
WT +1234.71 g
WT +1234.64 g
END
TIME 12:34:56
SIGNATURE
CR LF CR
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah