7-3 Correcting the Internal Mass Value (Only for FZ-i series)
The FZ-i series balance can correct the internal mass value by basing on an external weight.
The corrected mass value is maintained in non-volatile memory even if the AC adapter is removed.
(The FX-i series balance does not have the internal mass value correction function.)
Select an appropriate weight as listed below:
Model
Usable calibration weight
Adjustable range
FZ-120
i
100 g, 50 g
FZ-200
i
200 g, 100 g, 50 g
FZ-300
i
300
g,
200 g, 100 g, 50 g
-0.050 g to +0.050 g
FZ-1200
i
1000 g, 500 g
FZ-2000
i
2000 g, 1000 g, 500 g
FZ-3000
i
3000
g,
2000 g, 1000 g, 500 g
-0.50 g to +0.50 g
The calibration weight in bold type: factory setting
The calibration weight value can be adjusted within the range above.
Operation
1 With factory setting, the balance can not correct the internal
mass value. Refer to “8. FUNCTION SWITCH AND
INITIALIZATION” and set the function and the internal mass
value correction function.
2 With the weighing mode, press and hold the SAMPLE key to
display ba5fnc .
3 Press the SAMPLE key several times until Cal5et is
displayed.
If Cal5et is not displayed, perform the step 1.
4 Press the PRINT key to display Cal 0 .
Refer to “7-4 Calibration Using an External Weight”, to perform
the calibration.
5 After calibration, the balance displays ready
?
.
After placing the small break breeze on the balance, press the
PRINT key.
6 The balance corrects the internal mass value automatically
after displaying Cal5et .
7 After correcting the internal mass value, the balance displays
C
Calin , and performs the calibration automatically using the
corrected internal mass value.
8 The
balance
displays ba5fnc after displaying end .
9 Press the CAL key to return the weighing mode. Confirm that the
internal mass value has been corrected properly by placing the
mass used to performed the correction, on the weighing pan.
If the internal mass value has not been corrected properly,
repeat the procedure, starting at step 2 and adjust the
correction value.