Of 300 and 400 mm, Using the device – FRITSCH ANALYSETTE 18 User Manual
Page 23

n
DIN 66 165, part 1 basics - 1987 version
n
DIN 66 165, part 2 procedure - 1987 version
The standard sheets can be purchased from Beuth-Verlag, Berlin.
5.3 The maximum particle size permitted for test sieves
To avoid damaging the sieve base, the particle dimensions in a batch
should not exceed 10 w
0.7
(w is mesh width in mm).
Example
Nominal mesh width, w
Maximum particle size
mm
approx. mm
0.045
1
0.25
4
1.00
10
4.00
25
5.4 Recommended bulk volume for test sieves with diameters of 300 and 400 mm
Bulk volume for test sieves with a diameter
of 300 mm
Nominal mesh width
w
Feeding volume
Maximum sieve res-
idue
µm
approx. cm
3
approx. cm
3
25
33
17
32
39
20
45
45
23
63
53
26
90
65
32
125
75
38
180
90
45
250
105
53
355
120
60
Using the device
- 23 -