5 ambient conditions, 6 electrical connection, 5 ambient conditions 4.6 electrical connection – FRITSCH ANALYSETTE 18 User Manual
Page 19: Installation

3.
Select an appropriate position for the wall bracket of the separate
control device provided, and secure the bracket in place.
4.
Connect the controller provided for the sieve shaker to the device
drive.
4.5 Ambient conditions
WARNING!
Mains voltage!
– The device may only be operated indoors.
– The surrounding air may not carry any electrically con-
ductive dust.
– Maximum relative humidity 80% for temperatures up
to 31°C, linearly decreasing down to 50% relative
humidity at 40°C.
n
The room temperature has to stay between 5 - 40°C.
n
Altitudes up to 2000 m
n
Degree of pollution 2 according to IEC 664.
4.6 Electrical connection
DANGER!
Provide short-circuit protection!
Risk of damage due to short-circuits.
– Make sure that the socket is connected to a mains line
protected with a residual current circuit breaker.
DANGER!
Mains voltage!
Changes to the connection line may only be made by a
qualified technician.
Installation