10.0 TECHNICAL DATA
Resolution of the display
0,1% moisture content
0,5°C temperature
Measuring range
3 up to 20 % moisture content
Operation temperature
0°C up to 40°C
Storage temperature
-20°C up to 60°C
Temperature compensation
Automatically
Power supply
4 pcs. 1.5 Volt AA Alkaline batteries
(1800 measurements)
Automatically turn Off
After 6 minutes
Current consumption
55 mA (with light)
Display
128 x 64 matrix display (lighted)
Dimensions
260 x 70 x 250 mm
Weight
app. 1.3 kg (with batteries)
Degree of protection
IP 40
Scope of supply
Plastic case
Digital scale 1000g
Measuring cup 0.5 litre
4 x 1.5Volt AA Alkaline Batteries
Optional accessories
PC-Interface
Software LogMemorizer
Printer
Electromatic Equipment Co., Inc.
-5-
3.0
CALIBRATION CURVES
Name of Calibration
curve
Fill quantity
Measuring range
limits
280g wood pellet
280g
3 up to 20%
250g spec.pell. 1
250g
3 up to 20%
250g spec.pell. 2
250g
3 up to 20%
250g spec.pell. 3
250g
3 up to 20%
reference
--------
--------
Explanation of the curves:
280g wood pellet: To analyze the moisture of Ø6 and Ø8 mm wood pellets with
a maximum length of 30mm.
According to norm CEN/TS 14961
If you do not want to measure wood pellets (or wood pellets with smaller bulk
density), you have to determine the moisture according to CEN/TS 14774 once
per sort and assign these to a sort calibration curve (special pellets 1 to 3). Fill
up the BP1 and switch through all curves. The characteristic with the smallest
deviation is the correct one for this sort of pellets!
Filling quantity 250g
Reference: To test the humimeter BP1. Do not use for measuring the moisture!
List of calibration curves
Press in the measuring window the
or
key
for at least 3 seconds and a list with all available
sorts will appear. Select your sort by pressing
or
and confirm it with the
key. The
measurement will continue automatically.