Checkline ZP Recorder User Manual
Page 10
page 15
page 10
4. Comparator Tab
(optional key
)
14
The Comparator allows you to enter high/low force limits when
recording. High value equals red line; low value equals blue line
on the graph. These values correspond to Go/NoGo LEDs on the
force gauge when testing.
5. Preview Mode
(key )
Use these start and stop controls to quickly view force
profiles or confirm basic configuration parameters.
Statistics or file saving are not enabled when using preview
mode.
5
3. Axes Tab
(required key
)
13
Enter data or use the drop down list when entering axis values.
X-Axis: (Seconds)
0.5 – 10,000
Y-Axis: (Units)
0.001 – 2,000
Mouse Over Graph
The value for any data point (sample) on the graph can be dis-
played by positioning the mouse pointer over that specific point.