beautypg.com

Checkline FGE-XY User Manual

Page 5

background image

12. If calibration was unsuccessful, the display will show ERR. Re-
move the calibration weight, then press UNIT. Repeat the proce-
dure to attempt for a correct calibration.

TROUBLESHOOTING
The following are general checkpoints. Please call Shimpo or your
dealer for further assistance.

Force Gauges does not turn on: Check that power adapter is con-
nected properly to unit and outlet.

Error codes are displayed: Turn unit off, then back on to reset. If
error codes continue, see table below

CALIBRATION REqUIREMENTS
1. Secure and mount force gauge to a calibration stand and mount
the gauge upside down.

2. Obtain the appropriate full scale Metric range calibration weight
of your unit.

3. Turn POWER OFF

4. Attach the fixture on the shaft of the force gauge. Thread and
tighten onto shaft by hand only.

5. Press and hold the UNIT, PEAK and ZERO switches.

6. Press and release the POWER switch while still holding the
other 3 switches. Continue holding the 3 switches until the display
reads CAL, now release the three switches.

7. Press the UNIT switch. The display will blink SCN for 10 sec-
onds then show ZER. The force gauge is now ready for zero point
calibration.

8. Press ZERO to zero calibrate. Wait approximately 18 seconds.
The display will blink SCN then PEK. Do not press any other
switches or move the sensing shaft during calibration.

9. Hang the calibration weight on the hook and stabilize. The pro-
cess values will change. The gauge is now ready for full-scale
calibration.

10. Press the PEAK switch to begin full scale calibration. Display
blinks SCN. Do not press any keys or touch the unit or weight
during calibration. After approximately 18 seconds, the display
blinks END. After an additional 5 seconds, the display blinks OK
if successful.

11. Press the UNIT button, the unit automatically powers off.

5

CAL

-00-

“SCN”

-00-

ZER

88888

Err

88888

“SCN”

88888

“End”

88888

“oK”

88888

Small Display

Condition

Action

The load exceeds
120% of its capacity

Remove excessive
load.

EEPROM reading
error

Turn off, then turn on
again, if the display
does not return to
normal operation
send unit in for repair.

EEPROM writing error

OVR

ERR

3

ERR

4