Introduction, Overview – Cannon Instrument CT-1000HT (High Temperature Contant Temperature Bath) User Manual
Page 3

1
CANNON
®
CT-1000HT Constant Temperature Bath
Revision 1.1b—June, 2011;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
CHAPTER
1
INTRODUCTION
Overview
CT-1000HT function
The
CANNON
®
CT-1000HT Constant Temperature Bath is designed to
maintain precise temperatures at a wide range of settings for accurate
viscosity measurements. Because of its temperature stability and ease of
use, it is also suitable for any other application where temperatures must
be maintained within close tolerances.
Temperature range/stability
The
CANNON
®
CT-1000HT will maintain temperatures of 25°C to
100°C within 0.01°C, and temperatures of 101°C to 200°C within
0.03°C. The temperature stability of the CT-1000HT meets the accuracy
requirements of ASTM D 445 for kinematic viscosity measurements.
A built-in cooling coil, when connected to tap water or a cooling system,
permits operation below or slightly above ambient temperature. (Tem-
perature stability at or below ambient may vary depending on the quality
of the external cooling unit.)
Figure 1: The CT-1000