beautypg.com

Maintenance & calibration – Cannon Instrument SimpleVIS User Manual

Page 9

background image

The SimpleVIS

®

Viscometer was designed to require minimal maintenance

through simplicity of design and built-in electronic diagnostics. All new

instruments are shipped with factory calibrations. However, like most

metrological equipment, periodic maintenance and calibration is critical

to maintaining accurate long-term measurements. CANNON

®

Instrument

Company recommends the following intervals for maintaining the proper and

accurate operation of the SimpleVIS

®

Viscometer.

Basic Maintenance

The main chassis of the SimpleVIS

®

Viscometer does not contain any

user-serviceable parts and should be returned to the manufacturer (or an

authorized dealer service center) for repair when required. Basic instrument

maintenance should be carried out weekly or each time the effluent waste

bottle is emptied.
Cleaning the Viscometer Tube
To clean the viscometer tube simply start a wash cycle by tapping the “Ready

to Test” caption field in the Ready to Test Screen three times within two

seconds. This will cause the display to switch to the Wash Screen and start a

normal wash. See discretionary washing on page 12.
Changing the Tip on the Pipettor
Refer to the pipette manual for detailed information on using the device. The

tips can be wiped off after each test and reused. Use a clean dry paper towel

to wipe the tip off.
Cleaning the Display Screen
Turn off the SimpleVIS

®

Viscometer. Wet a soft, lint-free or micro fiber cloth

with clean or distilled water. Wring out as much water as possible to ensure

the cloth is damp but not wet. Wipe the screen in a gentle motion to remove

dust, oil, or fingerprint smudges. Finish cleaning the touch screen with a dry

lint-free cloth to wipe any excess moisture, and then turn the instrument back

on.

Maintenance & Calibration

9