Running a test – Cannon Instrument SimpleVIS User Manual

Page 6

background image

Injecting a sample

1. Using the supplied piston

operated pipettor, insert the tip

into a vial of test sample and

pull the measured amount

into the pipettor. Be sure no

bubbles exist in the oil or

results may be erratic. Refer

to pipettor instruction guide

for more information. The

pipettor volume setting should

be set to 450 ~ 500 μL

(0.45 - 0.50mL)

2. Place the pipettor at a 45°

angle into the sample/ solvent

injection port, located on

top of the instrument, and gently release the fluid in one easy operation

so that the entire sample is smoothly dispensed into the tube without

introducing bubbles.

The Ready to Test Screen contains a START button. Pressing the button is

optional as the injected sample will automatically be detected. The button is

useful for higher viscosity fluids that might take a while to trip the top sensor.
The run status indicator lamp will turn off during testing. When the viscosity

measurement is complete, the run status indicator lamp will blink and the

viscosity result will be displayed.

Running a Test