miniAV-X Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.Ob— March, 2012;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
62
To calibrate for temperature, see Chapter 3 for details. Temperature
calibration requires a reference thermometer with an accuracy meeting
the precision specification of ASTM D445.
To zero temperature offsets, click the
Service/Clear Temperature
Offsets
menu option for the desired instrument. The
Zero Temperature
Offsets
window will open. Click the check box to select each desired
action. Then click
OK
to clear the selected settings.
If the
Clear Current Offset
option is checked, VISCPRO will clear the
calibration point for the current bath temperature.