Instruction & Operation Manual

Version 1.0— February, 2008;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

4

Check the thermometer in the temperature bath to make sure the bath
is holding the proper temperature. If necessary, calibrate the miniAV-
X temperature control probe using the temperature calibration
procedure in Chapter 3.

5. Open the

View Instrument Group

window (if it is not already open)

by clicking

View Instrument

from the

Main

menu and selecting the

desired instrument group from the list box (see Chapter 4 for more
information).

The View Instrument Group window

6. Select the desired group ID and click

OK

to display the

Sample

Input View.

The Sample Input window (miniAV-X four--bath system)