BUCHI NIRFlex N-500 User Manual
Page 72
7 Maintenance and repairs
72
NIRFlex N-500 Operation Manual, Version F
Workflow to reset the lifetime counter of the primary lamp module
To establish the standard working conditions of the NIRFlex N-500:
a
b
c
Start the Operator software
•
Switch from “secondary lamp”
•
a to “primary lamp”
Resetting the hour
•
b counter will bring up a prompt for confirmation
Confirm the message
•
c to finally set the lamp lifetime counter to zero
➡ After this procedure, reboot the NIRFlex N-500 and restart the Operator software.