3 Technical data

15

NIRFlex N-500 Operation Manual, Version F

Technical data NIRFlex Solids Transmittance (cont.)
Photometric linearity

The “addition of filter technique” has been used.
A 2% transmission filter and a wavelength standard
(Rare Earth Oxide mixture) were used for the addition.
Both filters were measured individually and together in
series. The addition of the individual measurements has
been compared with the measurement of both filters ihn
series. The difference had been < 2 x 10

-7

T

at 7’876 cm

-1

Typical signal-to-noise ratio

open beam

5 mm white standard

rms for spectral segments of 300 cm

-1

in the range of

11’000 - 6’500 cm

-1

mean 2 x 10

-5

AU (16 scans; Blackman apodization)

mean 10 x 10

-5

AU (64 scans; Blackman apodization)

3 .3 .3

NIRFlex Liquids

Technical data NIRFlex Liquids
Detector

extended range InGaAs (temperature controlled)

Electric power supply provided from NIRFlex N-500

100−230 VAC ± 10 %, 50/60 Hz, 250 W

Ambient conditions

< 80 % relative humidity for T < 31 °C, linear
decreasing to 67 % at 35 °C, max. 2500 m, for indoor
use only

Ambient temperature

5−35 °C ( 25 ± 5 °C recommended)

Pollution degree

2

Installation category

II

Sample temperature range

Ambient temperature plus 10 °C up to 65 °C

Reproducibility of set sample temperature

±

0.5 °C

Temperature overshoot

< 5 °C

Overheating protection, automatic switch-off

T > 90 °C

Diameter of measurement spot

2 mm

Type of cuvettes to be used

Standard cuvettes 12.5 x 12.5 x 45 mm with path-
length of 2 mm, when spacers are used, 1, 5, 10 mm
cuvettes can be inserted as well

Time needed to reach a stable control of the set temperature Ambient temperature to 65 °C: 15 min