3 after the extraction process, 4 emptying the solvent tank, 5 reusing the solvent – BUCHI Extraction Unit E-816 SOX User Manual
Page 41
![background image](/manuals/656091/41/background.png)
6 Operation
41
E-812/816 Soxhlet Operation Manual, Version B
6. Place the beakers on the corresponding positions directly on the heating plate.
7. Lower the rack.
8. Activate the occupied positions.
9. Select a previous program or enter a new one.
10. Open the cooling water flow or switch on the connected chiller.
11. Start the extraction process.
6 .4 .3
After the extraction process
The end of an extraction process is indicated on the display and audible by a beep sound (if selected
in the instrument settings).
After an extraction process is finished, proceed as follows:
• Confirm the message on the screen by pressing the backward button.
• Empty the solvent tank.
• Move up the rack.
• Carefully remove the glass tubes or thimbles from the beakers by means of the pliers.
• Place the glass sample tubes or thimbles on a tissue and let them dry under the fume hood.
• Dry the beakers in a drying oven at 100 °C or in a vacuum oven at 200 mbar and 70 °C until they
have a constant weight.
• Let the beakers cool down to ambient temperature in a desiccator.
• Weigh the beakers and calculate the resulting fat content.
6 .4 .4
Emptying the solvent tank
• To empty the solvent tank put the solvent outlet hose into the waste container.
• Press Solv. and confirm the appearing message by pressing Yes.
NOTE
Make sure to empty the solvent tank after each extraction.
6 .4 .5
Reusing the solvent
• The solvent can be reused directly or after being distilled in a rotary evaporator.
When the instrument was used for the first time:
• Waste the solvent
When a different solvent was used before
• Waste the solvent
After incorrect procedure:
• Clean the glass assembly
• Clean the solvent tank
• Waste the solvent
NOTE
When working with petroleum ether from 40 - 60 °C: Renew the whole solvent after 10 extractions
to achieve a reproducible extraction behaviour. The mixture changes within time because the light
fraction evaporates and the heavy fraction accumulates.