9 skip function, 10 service functions, 9 skip function 6.3.10 service functions – BUCHI Extraction Unit E-816 SOX User Manual
Page 36
6 Operation
36
E-812/816 Soxhlet Operation Manual, Version B
6 .3 .9
Skip function
During the extraction process it is possible to skip one step and go directly to the next step, e.g. from
the extraction step to the rinse step. For this purpose, press Skip and confirm the appearing message
by pressing Yes.
Fig. 6.17: Skip
6 .3 .10
Service functions
In the main menu the submenu Service functions is available. Here you have an overview over all
sensors, functional efficiency of the valves, heating elements, information on the extraction unit and
the operating hours. Part of this information (all testing functions) is only available in the Service mode.
To change to the Service mode proceed as follows:
In the main menu select Mode, turn the selection knob to Service and press OK.
Fig. 6.18: Go to the Service Mode
Now go to the Service functions with the move up button and press the forward button. The following
screen appears: