Syncore

®

Accessories

version E

33

Maintenance of the intert gas module

You can use organic solvents like ethanol or acetone to
remove organic residues. You can use mild detergents and
a soft (plastic) brush to clean the module.

Punctured septa a can be removed by loosening the

screws b, and individually replaced by new septa. You

should visually inspect the septa after every reaction for
damages and contamination.

The inert gas module a can be removed by loosening the

two screws b. The inside of the vacuum tube can then be

cleaned with an organic solvent such as ethanol or acetone
or a mild detergent in water. Use a soft (plastic) brush.

You should visually examine the poly ethylene sealing discs
a after every reaction for contamination and wear. The

polyethylene sealing discs and the supporting rings b can

be simply replaced by loosening the PEEK screws c.

Cleaning is done by washing with alcohol or water.

Figure 40: Replacing septa

Figure 41: Taking apart an inert gas cover

Figure 42: Sealing system of the inert gas cover

a

b

a

b

a

b

c