BUCHI Syncor Polyvap User Manual
Page 15
Syncore
®
Accessories
version E
13
3.1 Syncore Product Line Accessories
The Syncore product line is modular throughout. Every ac-
cessory is usable at any time with a Syncore single platform
and with all standard configurations.
3.2 The condenser unit
Introduction
The condenser unit is used to condense out solvent va-
pors under a reduced pressure and to collect them in a
receiving flask. Cooling liquids may, for example, be water
at room temperature or liquids actively cooled in a cooling
aggregate.
During operation, the condenser unit is connected to a
vacuum source. Never use glassware if it is found to be
cracked. No manipulations may be made on either the
condenser unit or the receiving flask during operation.
Glass parts subjected to a vacuum may implode and cause
material damage and personal injuries. Ground glass joints
may be turned and the unit as a whole moved only has
been fully vented.
Attachment of the condenser unit S and C
• Screw the support rail a into the hole provided either on
the right or on the left of the back legs (depending on the
workplace and on the working procedure involved).
• Fasten the the pivoting clamp c with the cross-shaped
sleeve b to the stand.
• Fasten the cooler to the pivoting clamp.
• Using a clamp e, fasten the adapter d to the cooler.
• Likewise fasten the receiving flask f to the adapter with
a clamp g.
The firmly-secured hoses must not come into contact with
the heating plate or the heated and cooled racks or vacuum
connections and must be fixed by a clip.
3 Function,putting into operation, operation and maintenace
Figure 15: Attachment of the Condenser Unit
3 Function, putting into operation
operation and maintenance
a
c
e
d
f
b