Page 11

background image

3.12

Controller Display Messages

Display Description

Action

Required

….

Standby mode

Normal — Indicates that the Circuit Breaker/Power Switch is ON
and the Controller Power Switch is OFF

tx.xx

Power up self-test

Normal — Appears momentarily at startup

oCx.x

Calibration offset value

Normal — Current calibration offset value; refer to section 4.1

Hxxx

High Limit value

Normal — Current High Limit value; refer to section 3.6

Axxx

Auto-Refrigeration set point value

Normal — Current Auto-Refrigeration set point; refer to section
3.10 (Appears only on Refrigerating/Heating Circulators)

E-H1

High Limit set point too low

Error — The value entered is below the control temperature set
point. Refer to Setting High Limit, section 3.6

FLt 1

High Limit value exceeded

Error — Set a High Limit value higher, then turn main power to
the unit OFF and back ON; refer to section 3.6

FLt 2

EEPROM reset

Error — Turn Circuit Breaker/Power Switch OFF, hold P3 Button,
and then turn the Circuit Breaker/Power Switch back ON

FLt 3

Safety Set temperature exceeded

Error — Turn Circuit Breaker/Power Switch OFF, press Safety
Set Reset Button, and then turn the Circuit Breaker/Power Switch
back ON

FLt 4

Heating Triac failure

Error — Service required

FLt 5

Probe failure

Error — Service required

FLt 6

i2c error — Communication failure to
modulation board

Error — Service required (Appears only on Refrigerating/Heating
Circulators)



Section 4 - Calibration and Maintenance

4.1 Calibration

Calibration allows the user to match the Controller’s bath temperature display to an external certified
reference thermometer. Calibration is performed as follows:

Set the desired operating fluid temperature set point and allow temperature to stabilize.

Press the P2 and P3 simultaneously and release and repeat until the display reads (oCx.x). Press
P1 and hold until (Cal) is displayed. This will take about 2 seconds.

At one second intervals, the displayed value will alternate between the actual bath fluid temperature
and the current offset value, which is the difference between the factory calibration setting and the
user’s reference temperature sensor. The maximum offset is

±0.9°C from factory calibration.

To change the calibration offset value, rotate the Select/Set Knob until the display matches the
reading on the reference temperature sensor. The display will continue to alternate between the
offset value and the calibrated display temperature.

Press the Select/Set Knob or the P1 Button to accept the entered value. When the new calibration is
stored and the mode is exited, (

) will appear on the display.

NOTE: The displayed offset value will also be accepted if there is no activity for 20 seconds.

4.2 Heater

The heater should be kept clean. If deposits build up on the heater, they may be removed by
scrubbing with a non-metallic (plastic) abrasive pad. Do not use steel wool.