I.2 utilities, I.3 units, I.4 specifications – Brookfield LFRA Texture Analyzer User Manual
Page 4
Brookfield Engineering Labs., Inc.
Page
Manual No. M/05-370-B0507
I. Utilities
Input Voltage:
115 VAC or 230 VAC
Input Frequency:
50/60 Hz
Power Consumption:
150 VA
Fuse:
Two 4 amp, 5 x 20mm, Time-lag
Power Cord Color Code:
.
United.States.
Outside.United.States
Hot (live)
Black
Brown
Neutral
White
Blue
Ground (earth)
Green
Green/Yellow
Main supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
I.3 Units
The LFRA Texture Analyzer uses the SI system of units for all parameters.
Parameter
Unit
Abbreviation
Load
Grams
g
Distance
Millimeter
mm
Time
Seconds
s
Speed
Millimeter per second
mm/s
I. Specifications
Load:
Range
Resolution
Model
Grams
Grams
Accuracy*
LFRA-100
0-100
0.01
±0.5%FSR
LFRA-1000
0-1000
0.10
±0.5%FSR
LFRA-1500
0-1500
0.20
±0.5%FSR
LFRA-4500
0-4500
0.50
±0.5%FSR
FSR = Full Scale Range
*This value applies to the complete range of operating temperatures as defined in Section I.4
(Environmental Conditions). This value will be improved if the environment is held to a
stable temperature.
Trigger Point:
Range
Resolution
Model
Grams
Grams
LFRA-100
0.1-10
0.01
LFRA-1000
0.2-100
0.10
LFRA-1500
0.2-150
0.10
LFRA-4500
0.6-500
0.10
Speed: 0.1 to 1.0 mm/s in increments of 0.1 mm/s
1 to 10 mm/s in increments of 1 mm/s
Accuracy: ±0.1% of setpoint