III.2

Signal Output

The signal output option includes parallel printer interface and analog voltage using a 25 pin, Type
D (male) connector (Brookfield Part No. KU-1C) as follows:

0 - 5 Volt DC Analog Voltage:

Gram output:

4 mV/gram

KU output:

20 mV/KU

Range, KU:

0 - 144

Range, gm:

0 - 1,095

Resolution, KU:

1

Resolution, gm:

5

Interface pin outs:

pin 18 voltage
pin 19 ground (earth)

Centronics Parallel Printer Output

Pin

Function

1

STB

2

D0

3

D1

4

D2

5

D3

6

D4

7

D5

8

D6

9

D7

25

GND

The printer will print the Krebs Units or grams depending on the position of the selector switch (e.g. “96
KU” or “745 gm”). A series of dashes (“- - - KU”) indicates an overrange condition.

The output connector is located on the left side of the viscometer cover.

III.3 Paste Spindle

The paste spindle is similar to the Krebs paddle-type spindle. The design consists
of offset rod-type vanes, approximately 2 mm diameter x 19 mm long. This
spindle is suitable for use with high consistency materials such as roller mill
pastes. Do not use this spindle to report normal Krebs Unit measurements.

Paste Spindle