Brookfield KU-1 Viscometer User Manual
Page 9
-9-
III.2
Signal Output
The signal output option includes parallel printer interface and analog voltage using a 25 pin, Type
D (male) connector (Brookfield Part No. KU-1C) as follows:
0 - 5 Volt DC Analog Voltage:
Gram output:
4 mV/gram
KU output:
20 mV/KU
Range, KU:
0 - 144
Range, gm:
0 - 1,095
Resolution, KU:
1
Resolution, gm:
5
Interface pin outs:
pin 18 voltage
pin 19 ground (earth)
Centronics Parallel Printer Output
Pin
Function
1
STB
2
D0
3
D1
4
D2
5
D3
6
D4
7
D5
8
D6
9
D7
25
GND
The printer will print the Krebs Units or grams depending on the position of the selector switch (e.g. “96
KU” or “745 gm”). A series of dashes (“- - - KU”) indicates an overrange condition.
The output connector is located on the left side of the viscometer cover.
III.3 Paste Spindle
The paste spindle is similar to the Krebs paddle-type spindle. The design consists
of offset rod-type vanes, approximately 2 mm diameter x 19 mm long. This
spindle is suitable for use with high consistency materials such as roller mill
pastes. Do not use this spindle to report normal Krebs Unit measurements.
Paste Spindle