Storage, Troubleshooting & replacement parts, Operator’s manual – ANKOM 200 Series User Manual
Page 30

Operator’s Manual
pg. 30
Rev C 1/14/15
Check the Bag Suspender
The Bag Suspender should be checked every three to six months or if fiber values are higher than normal or
inconsistent. To check the Bag Suspender, follow the procedure detailed below.
1. Check the trays for melting.
The pictures shown are examples of extreme
cases. However, for proper operation you
must replace trays that show signs of melting
or wear.
2. Check that the bottom tray is flat.
If the bottom tray is concave (see picture) the
bag suspender will catch in the vessel and
melt.
3. Check the tip for excessive wear.
Replace worn tips (see pictures).
Storage
Acid residue can corrode the instrument if left on or in the instrument. Before storing the instrument (or leaving the
instrument unused) for greater than one month, rinse the Vessel thoroughly with water using the following
procedure.
1.
Fill the Vessel with water
.
2.
Press the
A
A
G
G
I
I
T
T
A
A
T
T
E
E
button so that it is lit.
3.
After 10 minutes, press the
A
A
G
G
I
I
T
T
A
A
T
T
E
E
button so that it is NOT lit. This will turn the agitation off.
4.
Open the Exhaust Valve and exhaust the hot water.
5.
After the water has been exhausted, close the Exhaust Valve.
Troubleshooting & Replacement Parts
The ANKOM technology web site has the most current troubleshooting and replacement parts information.
Therefore, if you have any questions about the operation of your ANKOM
200
Fiber Analyzer, or if you need
replacement parts, please visit our web site at
www.ankom.com.
Melted trays
Good
Tip
Worn
Tip