Quick-Check User Instructions

23

7 Troubleshooting

Problem

Possible Cause

Solution

Powers on momentarily and
turns off

Low battery

Replace with high quality alkaline batteries

Does not power on

Low battery

Replace with high quality alkaline batteries

Batteries installed
backwards or no
spring contact

Insure that positive terminals of both batteries
(nub) face inward – towards the black cap. Check
that spring is attached to the battery cap.

Software reset

Remove battery cap & reinstall after one minute.
Attempt to turn power on again.

Display contrast too
light

Hold the Right Arrow key down while pressing
the F2 key several times to increase the display
contrast. If nothing occurs, release both keys.
Press the power button and try again.

Display is completely dark

Display contrast too
dark

Hold the Arrow key down while pressing the F1
key several times to decrease the display
contrast.

Display drifts downward
once installed

Wire material is
creeping and internal
friction between wires
is relieved.

This is normal behavior of wire. Lower display
resolution to mask this effect.

Temperature not accurate

Instrument changed
temperature
environments

Allow instrument to remain in environment until
temp stabilizes or enter temp manually

Instrument exposed
to sun

Enter temp manually