Quick-Check User Instructions
There is a keypad shortcut for increasing and decreasing contrast.
While in normal display mode press the Arrow key and the 2nd
softkey simultaneously to increase contrast. Press the Arrow key
and first softkey simultaneously to decrease contrast.
Blite - Not currently used.
About Press this softkey to see the following information:
Device - Press this softkey to show a list of information about the
Quick-Check; serial number, capacity rating, hardware and
software revision levels. Press any key to return to the previous
softkey set.
Calib - Press this softkey to show Calibration Points and the
calibration information for the current wire size. Follow the on-
screen prompts.
O. Load - Press this softkey to show an audit count of the number
of times the unit has been overloaded beyond 125% of capacity.
Press any key to return to the previous softkey set.
Zero - Press this softkey to show the deadload analysis of the
Quick-Check. Press any key to return to the previous softkey set.
Ntwrk - Not used in the Quick-Check.
Test
Press this softkey and the following softkeys appear:
Batt - Press this softkey to test the battery level.
A-D - Press this softkey to display the A to D counts.
Disp. - Press this softkey to perform a test of the display pixels.
Keys - Press this softkey to test the keypad.
Comm - Not used in the Quick-Check.
Ntwrk - Not used in the Quick-Check.
Config
This is a password protected menu. See Configuration Mode
.
Press the ESC key to return to the normal operating mode. If you made changes to the
configuration of the unit, you are prompted to save them or abort the changes. Do so
and the unit returns to normal operation mode.