GS series User’s Manual REV 2.1 22/01/12

14


A total of overload count

% offset of last calibrate


% current offset

Figure 8 Diagnostic Menu

If the % offset is between 5% - 10 % please contact your

supplier to arrange a recalibration of your gauge.

If the % offset is greater than 10% please contact your supplier

to arrange for load cell replacement.

These values are given as an indicator only – the need for

calibration/repair may vary according to the individual characteristics of
the load cell.

6) ABOUT This shows the information of your gauge (Firmware
revision, Model, Capacity, Serial number). To access ABOUT menu, go to

main menu page and press UP and DOWN to move the cursor point to

ABOUT and press ENTER key the display will show About menu page.
Press ESC key to return to main menu page.

ABOUT

FIRMWARE REV. : 1.00
MODEL: GS
CAPACITY: 100 N

S/N: 05130001

Figure 9 About Menu




DIAGNOSTIC


OVERLOAD COUNT: 2


ORG. OFFSET : +0.4 %
CUR. OFFSET : +0.4 %