A total of overload count
% offset of last calibrate
% current offset
Figure 8 Diagnostic Menu
If the % offset is between 5% - 10 % please contact your
supplier to arrange a recalibration of your gauge.
If the % offset is greater than 10% please contact your supplier
to arrange for load cell replacement.
These values are given as an indicator only – the need for
calibration/repair may vary according to the individual characteristics of
the load cell.
6) ABOUT This shows the information of your gauge (Firmware
revision, Model, Capacity, Serial number). To access ABOUT menu, go to
main menu page and press UP and DOWN to move the cursor point to
ABOUT and press ENTER key the display will show About menu page.
Press ESC key to return to main menu page.
ABOUT
FIRMWARE REV. : 1.00
MODEL: GS
CAPACITY: 100 N
S/N: 05130001
Figure 9 About Menu
DIAGNOSTIC
OVERLOAD COUNT: 2
ORG. OFFSET : +0.4 %
CUR. OFFSET : +0.4 %