Dillon GS Plus Force Gauge User Manual
Page 11
GS series User’s Manual REV 2.1 22/01/12
11
2) PASS-FAIL the Pass-Fail feature used to set a defined acceptable
maximum and minimum forces gap for measuring. It activate by setting
the lower level and upper level forces limit If the forces value is within
the gap level, the display will show message PASS. Any reading values
outside this gap (higher or lower), the display will show message FAIL. If
you activate this feature the PF symbol will display on main display.
To access PASS-FAIL menu, Press UP and DOWN to move the
cursor point to PASS-FAIL and press the ENTER key the display will show
the Pass-Fail menu page. Press ESC key to return the main menu page.
Figure 6 Pass-Fail Menu
Use LEFT ARROW keys to move cursor point to the desired value.
Use UP and DOWN keys to change the value, press and hold to scroll
values. Use RIGHT ARROW key to change the unit. Press ENTER to save
setting and return to main menu page.
*Pass-Fail feature will automatically disabled if you set LOWER
and UPPER = 0 N.
*LOWER must be less than the UPPER.
example LOWER LEVEL = 0 N, UPPER LEVEL = 20 N
Load
The “UPPER” LED will ON. Another LED OFF. Analog Out = 0V
Upper level
The “OK” LED will ON. Another LED OFF
Time
Figure 6a
PASS FAIL MENU
UPPER = 2.5 N
LOWER = 1.0 N
Press ‘Zero’ key to
Clear both value.