beautypg.com

Technical data, 1 general technical data – Seca 745 band_1 User Manual

Page 42

background image

42 •

10.2 Information on servicing and subsequent verification

(seca 745)

We recommend having your device serviced prior to
subsequent verification.

NOTICE!

Incorrect measurements as a result of poor
servicing

– Please only have servicing and repairs

performed by an authorized service partner.

– You can find service partners in your area at

www.seca.com or by sending an e-mail to
[email protected].

Have an authorized technician perform subsequent
verification according to national legal regulations. The
year of first verification can be found after the CE sym-
bol on the rating plate above the number of the notified
body 0109 (Hessen Weights & Measures Office [Hes-
sische Eichdirektion, Germany]).

Subsequent verification is necessary whenever one or
more verification seals are damaged.

11.TECHNICAL DATA

11.1 General technical data

General technical data

seca 745/725

Dimensions

• Width
• Height
• Depth

555 mm
185 mm
320 mm

Weight approx.

6

kg

Ambient conditions, operation

• Temperature
• Air pressure
• Humidity

+10 °C to +40 °C (50 °F to 104 °F)

700 - 1060 hPa

30 % - 80 %, no condensation

Ambient conditions, storage

• Temperature
• Air pressure
• Humidity

-10 °C to +65 °C (14 °F to 149 °F)

700 - 1060 hPa

0 % - 95 %, no condensation