6-2-5 Perform the test described in paragraph 6-2-4 as

follows:

In the TEM cell, test at 50, 300 and 500 MHZ

In the anechoic chamber, test at 0.9, 1.2 and 2.5

GHz

6-2-6 From the data in paragraph 6-2-5, calculate the

average power density measured over the band

by taking the square root of the product of the

highest and lowest readings recorded. Record

the resultant on the data sheet for P

CAL

.

6-2-7 Divide 0.5 mW/cm

2

by the power density level

recorded on the data sheet for P

CAL

. This

determines the correction multiplier, M.

6-2-8 Alignment can now be performed at any one of

the calibration frequencies specified in step 6-2-

5. Select a convenient calibration frequency.

6-2-9 At the selected calibration frequency, multiply by

M the indicated power density from the data

recorded in paragraphs 6-2-4 and 6-2-5 above.

Record this as P

REF

.

6-2-10 Locate the calibration potentiometer access hole

behind the “Made in USA” label. Remove the

label only if adjustment is needed.

6-2-11 Apply the P

REF

power at the calibration frequency.

Adjust this potentiometer such that the alarm just

starts to indicate. Repeat this adjustment as

required.

6-2-12 This completes the RF calibration. Replace the

“Made in USA” label. Attach a new calibration