3-4-1 Set-Up Mode

The "Set-up" mode is used to set the alarm warning level

and to enable/disable the audible alarm. To enter this

mode, set the "ON-OFF" switch to ON. Then depress the

MODE switch after the initial three second period

referenced in paragraph 3-4. The LCD display will

indicate "S.U.". At this point, the alarm warning level

can be changed (paragraph 3-4-2).

3-4-2 Setting Alarm Warning Level

In the "Set-Up" mode (paragraph 3-4-1), depress and

hold the MODE switch. After three seconds, the preset

alarm warning level will be displayed. Release the MODE

switch. Depressing and holding the MODE switch again

at this point will result in a rapidly increasing alarm

warning level indication. Continue to depress the MODE

switch until the desired alarm level is approached. Then,

release the MODE switch and pulse it at a convenient

rate until the desired alarm warning level is reached. Once

it is reached, set the ON-OFF switch to OFF and then to

ON again. The alarm warning level that has just been set

will remain in memory until changed again as described

above.

3-4-3 Measurement Modes

The Model HI-3510 Measurement Mode can be set to

either "Instantaneous" or "Average" by depressing and

releasing the MODE switch while the unit is in its normal

operating mode. See paragraphs 3-4-3-1 and 3-4-3-2.