3-4-1 Set-Up Mode
The "Set-up" mode is used to set the alarm warning level
and to enable/disable the audible alarm. To enter this
mode, set the "ON-OFF" switch to ON. Then depress the
MODE switch after the initial three second period
referenced in paragraph 3-4. The LCD display will
indicate "S.U.". At this point, the alarm warning level
can be changed (paragraph 3-4-2).
3-4-2 Setting Alarm Warning Level
In the "Set-Up" mode (paragraph 3-4-1), depress and
hold the MODE switch. After three seconds, the preset
alarm warning level will be displayed. Release the MODE
switch. Depressing and holding the MODE switch again
at this point will result in a rapidly increasing alarm
warning level indication. Continue to depress the MODE
switch until the desired alarm level is approached. Then,
release the MODE switch and pulse it at a convenient
rate until the desired alarm warning level is reached. Once
it is reached, set the ON-OFF switch to OFF and then to
ON again. The alarm warning level that has just been set
will remain in memory until changed again as described
above.
3-4-3 Measurement Modes
The Model HI-3510 Measurement Mode can be set to
either "Instantaneous" or "Average" by depressing and
releasing the MODE switch while the unit is in its normal
operating mode. See paragraphs 3-4-3-1 and 3-4-3-2.