SECTION VI

CALIBRATION

6-1

GENERAL

The Model HI-3510 should be calibrated at 12 month

intervals. To do so requires the use of highly specialized

test equipment and facilities, such as a radio-frequency

anechoic chamber and TEM cells, wherein fields of

known power density can be accurately established. It

should also be noted that to establish the required power

densities will require CW sources of 10W or greater

together with accurately calibrated test antennas and

power monitoring equipment. If facilities of this type are

not available, the unit should be returned to Holaday

Industries or to another qualified calibration facility for

this service.

6-2

CALIBRATION PROCEDURE

6-2-1 Set the alarm warning level to 0.5 mW/cm

2

as

described in paragraph 3-4-2.

6-2-2 Disable the latch function as described in

paragraph 3-4-4.

6-2-3 Following the procedures described below,

position the unit in a 50 - 500 MHZ TEM cell and

in a 900 MHZ to 2.5 Ghz anechoic chamber.

The preferred orientation of the E-field is parallel

to the vertical axis of the unit.

6-2-4 At each test frequency, raise the power density

level at the unit until an alarm is indicated.

Record this power density level on the test data

sheet in mW/cm

2

.

Archived 4/2/10