Archived 4/2/10 – ETS-Lindgren HI-3510 RF Radiation Badge Broadband Electromagnetic Radiation Detector (Archived) User Manual

Page 17

background image

Page — 12

HI

n

3510 Manual

3-4-3-1

“Average” Measurement Mode

When the AVG annunciator on the left side of the display

area is visible, a six-minute running average is maintained

by the unit. Power density is sampled every 3.6

seconds. During the first six minutes (i.e.-360 seconds

or 100 measurements), the running average of all

measurements taken will be displayed. After six minutes,

each "new" measurement will replace the measurement

taken 363.6 seconds prior to that "new" measurement.

Therefore, the average of the most recent 100

measurements will always be displayed.

The average of the latest 100 measurements is also

compared by the unit to the alarm warning level set by

the operator (paragraph 3-4-2) to determine the %

ALARM LEVEL ratio. (During the first six minutes of

operation, the running average of all measurements taken

will be compared by the unit to the alarm warning level

set by the operator to determine the % ALARM LEVEL

ratio.) If the % ALARM LEVEL ratio reaches 100%, the

unit's alarm circuitry will be activated, the red LED and

the ALARM annunciator on the LCD will flash every 3.6

seconds, and the audible alarm will sound every 3.6

seconds.

3-4-3-2

"Instantaneous" Measurement Mode

When the INST annunciator on the left side of the LCD

display area is visible, the measured power density is

sampled and displayed every 1.8 seconds. (Note that the

Model HI-3510 saves data for six minutes regardless of

the display selected. Switching between "Average" and

"Instantaneous" does not affect the data being stored.)