41

Probe

Command

Description

Probe Response

Rx

Set range.

x=1, 2, 3, 4, or N (next

range)

Rx, where x is the range

x=returns the range

currently in use

x=1, 2, 3, 4 enables the

selected range

x=N sets the Field Probe to

the next higher range

Tx

Read temperature.

x=C or F

For TF command:

Txxx

xxx=temperature in

degrees in Fahrenheit

For TC command:

Txxx

• xxx=temperature in

degrees in Centigrade

Ux

Set unit type.

x=1, 2, 3, or N (next

unit)

• 1=V/m

• 2=mW/cm²

• 3=[V/m]²

Z

Zero

Null

Send the ASCII null

character.

This is a special command that

can be used as the initial

command to the Field Probe

after it is turned on. The probe

responds with N.