Disassembly for method rs101 testing – ETS-Lindgren 7606 Coil Antenna/Shielded Coil User Manual
Page 29

If all instrumentation is operating properly and is accurately calibrated, readjust
the source to produce the correct output from the Model 7606. Use the ratio (or
difference in dB) between the initial setting of the source to the corrected setting
of the source as a correction factor in the later EMI susceptibility measurements.
Disassembly for Method RS101 Testing
Before disassembling Model 7605 and Model 7606, verify that the test signal
source is turned off or the output is reduced to zero.
1.
Remove the 7605/7606 assembly from the tripod.
2.
Unscrew the nylon bolt.
3.
Remove the Model 7606 from the Model 7605.
MIL-STD-462D Method RS101 Testing
| 29